If asked to recalibrate your O2 sensor by a support technician, please follow these instructions.
- Switch the unit off by touching the front switch off button.
- Let the system depressurize
- Remove the sensor cap (E) and expose the sensor to ambient air.
- Wait for 60 seconds.
- Enter the maintenance menu , screen 3/5 wait until the oxygen level reads 20.9%.
- Then go to screen 5/5 and touch O2 20.9% button. The system will ask for confirmation. Select YES.
- Reconnect the tube (F) and the sensor cap (E) with the sensor (C).
Please note:
The sample flow is factory preset by with FCV3 and normally should not be adjusted. Tampering with this valve will have a major affect on the operation of the generator. Please inform our service team if you suspect that the sample flow is incorrect.
O2 sensor Replacement
O2 sensors have a limited lifespan, usually around 3 years. Extended storage also affects operational lifespan. Contact us if you need to purchase a replacement sensor, which are normally held in stock in our warehouse.